Electronic caries monitor (ECM) device (ECM II / ECM III / ECM IV) vs Reference standard (histology, excavation, or enhanced visual examination)

Sensitivity · Children, adolescents, or adults presenting asymptomatically or suspected of early enamel caries (in vivo), or extracted teeth of…
1 finding from 1 review
  • Electrical conductance for the detection of dental caries 2021 · Cochrane All outcomes
    Up to 3 months between index test and reference standard (in vivo studies)
    ECM (4 studies, 475 tooth surfaces): sensitivity ranged from 0.61 to 0.98 (0.61-0.65 in three studies; 0.98 in the most recent ECM IV study). No evidence of a difference in accuracy by device type (Chi2=0.05, df=1,…
    4 studies
    GRADE not reported