Cochrane · DTA · CD014547

Electrical conductance for the detection of dental caries

Macey, R, Walsh, T, Riley, P, et al. · Cochrane Database of Systematic Reviews · 2021 · Cochrane Library

Oral HealthDental cariesAMSTAR 2: not assessed
Full textEvery question was extracted from the full review (methods and results), not only its Summary of Findings table.
Questions
15
current version
GRADE rated
3
No studies
4
Included studies
7
as reported
No studies 4 · 26.7%Ungraded 8 · 53.3%Very low 3 · 20%Low 0 · 0%Moderate 0 · 0%High 0 · 0%

Ungraded means an effect was reported without a GRADE rating; it is not a gap. No studies means the review found nothing to answer the question. The six categories are a display order, not a score.

Review question

ElementPlanned (methods)Found (results)
PopulationChildren; Adolescents (immature permanent dentition); Adults (mature permanent dentition); Asymptomatic patients presenting for routine examination; Suspected early/enamel caries; Extracted teeth (in vitro); Advanced dentine/cavitated lesions excludedPrimary dentition/children (n=3 studies); Permanent dentition/adults (n=4 studies); In vitro extracted teeth (n=3); In vivo clinical setting (n=4); countries Brazil/UK/Denmark/Turkey
InterventionElectrical conductance devices (alone, as adjunct to visual examination, or in combination); Electronic caries monitor (ECM); CarieScan Pro; Vanguard electronic caries detectorElectronic caries monitor (ECM) (n=4 studies, 475 tooth surfaces; ECM II/III/IV); CarieScan Pro (n=3 studies, 244 tooth surfaces)
ComparisonReference standards: histology; enhanced visual examination; radiographic examination; operative exploration/excavation; composite reference standard; and comparisons between the electrical conductance devicesHistology (n=6 studies); Enhanced visual examination (n=1 study)
OutcomesSensitivity; Specificity; Diagnostic odds ratio (DOR); SROC/HSROC summary curveSensitivity (n=7 studies, range 0.55-0.98); Specificity (n=7 studies, range 0-1.00); Diagnostic odds ratio (n=7 studies pooled, DOR 15.65, 95% CI 1.43-171.15)
Study designsDiagnostic test accuracy studies (prospective; single index test or direct comparison); RCTs of test accuracy; case-control accuracy studies (excluded from primary analysis); in vitro and in vivoDiagnostic test accuracy studies (n=7; 8 datasets); in vitro (n=3); in vivo (n=4)
TimingUp to 3 months between index test and reference standard (in vivo)Not specified

Search date: 2019-04. Databases: MEDLINE Ovid; Embase Ovid; US NIH Ongoing Trials Register (ClinicalTrials.gov); WHO International Clinical Trials Registry Platform; reference lists of included papers and previous systematic reviews

Questions and results

Electrical conductance devices (electronic caries monitor (ECM) and/or CarieScan Pro), used as an adjunct to clinical oral examination vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Children, adolescents, or adults presenting asymptomatically or suspected of early enamel caries (in vivo), or extracted teeth of children/adults (in vitro); target condition = non-cavitated coronal caries at the enamel threshold on occlusal surfaces. Studies intentionally including caries into dentine or frank cavitations were excluded.
7 questions
OutcomeEffectEvidenceStatus / GRADE
Sensitivity (test accuracy for detecting enamel caries)
Up to 3 months between index test and reference standard (in vivo studies)
SoF findings (HSROC; no summary operating point reported): 'ECM: sensitivity from 0.61 to 0.98 ... CarieScan Pro: sensitivity from 0.55 to 0.91'. Across all 7 studies (8 datasets), covering 719 tooth surfaces or sites…7 studiesVery low
Downgraded: risk of bias, imprecision, inconsistency
Specificity
Up to 3 months between index test and reference standard (in vivo studies) · Primary dentition
Primary dentition (3 studies): specificity ranged from 0 to 0.92.3 studiesUngraded
Sensitivity
Up to 3 months between index test and reference standard (in vivo studies) · Permanent dentition
Permanent dentition (4 studies): sensitivity ranged from 0.55 to 0.65.4 studiesUngraded
Specificity (test accuracy for detecting enamel caries)
Up to 3 months between index test and reference standard (in vivo studies)
SoF findings (HSROC; no summary operating point reported): 'ECM: specificity from 0.73 to 0.96 ... CarieScan Pro: specificity from 0 to 1.00'. Across all 7 studies (8 datasets), covering 719 tooth surfaces or sites (no…7 studiesVery low
Downgraded: risk of bias, imprecision, inconsistency
Specificity
Up to 3 months between index test and reference standard (in vivo studies) · Permanent dentition
Permanent dentition (4 studies): specificity ranged from 0.73 to 1.00.4 studiesUngraded
Diagnostic odds ratio (overall test accuracy)
Up to 3 months between index test and reference standard (in vivo studies)
Diagnostic odds ratio (DOR) 15.65 [1.43, 171.15]7 studiesVery low
Downgraded: risk of bias, imprecision, inconsistency
Sensitivity
Up to 3 months between index test and reference standard (in vivo studies) · Primary dentition
Primary dentition (3 studies): sensitivity ranged from 0.91 to 0.93. No evidence of a difference in accuracy by dentition (removing accuracy covariate: Chi2=3.15, df=1, P=0.08).3 studiesUngraded

Electronic caries monitor (ECM) device (ECM II / ECM III / ECM IV) vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Children, adolescents, or adults presenting asymptomatically or suspected of early enamel caries (in vivo), or extracted teeth of children/adults (in vitro); target condition = non-cavitated coronal caries at the enamel threshold on occlusal surfaces. Studies intentionally including caries into dentine or frank cavitations were excluded.
2 questions
OutcomeEffectEvidenceStatus / GRADE
Sensitivity
Up to 3 months between index test and reference standard (in vivo studies)
ECM (4 studies, 475 tooth surfaces): sensitivity ranged from 0.61 to 0.98 (0.61-0.65 in three studies; 0.98 in the most recent ECM IV study). No evidence of a difference in accuracy by device type (Chi2=0.05, df=1,…4 studiesUngraded
Specificity
Up to 3 months between index test and reference standard (in vivo studies)
ECM (4 studies, 475 tooth surfaces): specificity ranged from 0.73 to 0.96.4 studiesUngraded

CarieScan Pro device vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Children, adolescents, or adults presenting asymptomatically or suspected of early enamel caries (in vivo), or extracted teeth of children/adults (in vitro); target condition = non-cavitated coronal caries at the enamel threshold on occlusal surfaces. Studies intentionally including caries into dentine or frank cavitations were excluded.
2 questions
OutcomeEffectEvidenceStatus / GRADE
Sensitivity
Up to 3 months between index test and reference standard (in vivo studies)
CarieScan Pro (3 studies, 244 tooth surfaces): sensitivity ranged from 0.55 to 0.91. No evidence of a difference in accuracy by device type (Chi2=0.05, df=1, P=0.82).3 studiesUngraded
Specificity
Up to 3 months between index test and reference standard (in vivo studies)
CarieScan Pro (3 studies, 244 tooth surfaces): specificity ranged from 0 to 1.00.3 studiesUngraded

Electrical conductance devices (electronic caries monitor (ECM) and/or CarieScan Pro), used as an adjunct to clinical oral examination vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Children/adolescents/adults or extracted teeth; target condition = non-cavitated enamel caries on PROXIMAL (approximal) tooth surfaces.
1 question
OutcomeEffectEvidenceStatus / GRADE
Outcome NR
Up to 3 months between index test and reference standard (in vivo studies)
No eligible studies. All included studies assessed occlusal surfaces only ('All studies assessed the use of the electrical conductance devices on occlusal surfaces.'); proximal surfaces were a planned surface subgroup…0 studiesNo studies

Electrical conductance devices (electronic caries monitor (ECM) and/or CarieScan Pro), used as an adjunct to clinical oral examination vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Children/adolescents/adults or extracted teeth; target condition = non-cavitated enamel caries on SMOOTH tooth surfaces.
1 question
OutcomeEffectEvidenceStatus / GRADE
Outcome NR
Up to 3 months between index test and reference standard (in vivo studies)
No eligible studies. All included studies assessed occlusal surfaces only; smooth surfaces were a planned surface subgroup with no data.0 studiesNo studies

Electrical conductance devices (electronic caries monitor (ECM) and/or CarieScan Pro), used as an adjunct to clinical oral examination vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Teeth with previously applied restorations (secondary caries) or pit and fissure sealants; target condition = non-cavitated enamel caries.
1 question
OutcomeEffectEvidenceStatus / GRADE
Outcome NR
Up to 3 months between index test and reference standard (in vivo studies)
No eligible studies. 'No included studies reported inclusion criteria of previously applied restorations or pit and fissure sealants' (a planned subgroup).0 studiesNo studies

Electrical conductance devices (electronic caries monitor (ECM) and/or CarieScan Pro), used as an adjunct to clinical oral examination vs Reference standard: histology or enhanced visual examination (with or without radiographs)

Population: Samples with a low (<15%) prevalence of caries into dentine (d3/D3); target condition = non-cavitated enamel caries.
1 question
OutcomeEffectEvidenceStatus / GRADE
Outcome NR
Up to 3 months between index test and reference standard (in vivo studies)
No eligible studies. 'No studies reported the inclusion of a low prevalence (0% to 14%) of d3/D3 of tooth surfaces'; all included studies reported d3/D3 prevalence of 15% or above, so the planned low-prevalence…0 studiesNo studies