Laser fluorescence (DIAGNOdent 2190/pen) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a summary ROC AUC of 0.83 for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Conventional bitewing radiography (E-speed film) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, conventional E-speed bitewing radiography had a pooled diagnostic odds ratio of 31.40 (95% CI 14.31, 68.93) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Digital bitewing radiography, phosphor plate (PSP) vs Direct visual examination after tooth separation or bioptical cavity preparation

Diagnostic accuracy for proximal caries detection (dentin detection level) · Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)

Under in vivo conditions, phosphor-plate digital bitewing radiography had a pooled specificity of 0.60 (95% CI 0.20, 0.90) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Laser fluorescence (DIAGNOdent 2190/pen) vs Direct visual examination after tooth separation or bioptical cavity preparation

Diagnostic accuracy for proximal caries detection (dentin detection level) · Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)

Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled specificity of 0.60 (95% CI 0.15, 0.93) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Conventional bitewing radiography, F-speed film vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, conventional F-speed bitewing radiography had a pooled diagnostic odds ratio of 8.7 (95% CI 4.2, 18.1) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Conventional bitewing radiography (D-speed film) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, conventional D-speed bitewing radiography had a pooled specificity of 0.99 (95% CI 0.96, 0.999) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Digital bitewing radiography, sensor-based (CCD/CMOS) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, sensor-based digital bitewing radiography had a pooled diagnostic odds ratio of 12.84 (95% CI 6.56, 25.13) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Conventional bitewing radiography (E-speed film) vs Direct visual examination after tooth separation or bioptical cavity preparation

Diagnostic accuracy for proximal caries detection (dentin detection level) · Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)

Under in vivo conditions, conventional E-speed bitewing radiography had a pooled specificity of 0.91 (95% CI 0.80, 0.96) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Digital bitewing radiography, phosphor plate (PSP) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, phosphor-plate digital bitewing radiography had a summary ROC AUC of 0.92 for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Digital bitewing radiography, sensor-based (CCD/CMOS) vs Direct visual examination after tooth separation or bioptical cavity preparation

Diagnostic accuracy for proximal caries detection (dentin detection level) · Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)

Under in vivo conditions, sensor-based digital bitewing radiography had a pooled sensitivity of 0.96 (95% CI 0.91, 0.98) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Visual examination (VE) vs Direct visual examination after tooth separation or bioptical cavity preparation

Diagnostic accuracy for proximal caries detection (dentin detection level) · Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)

Under in vivo conditions, visual examination had a pooled specificity of 0.76 (95% CI 0.11, 0.99) for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Visual examination (VE) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, visual examination had a summary ROC AUC of 0.85 for proximal caries detection at the dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Conventional bitewing radiography, F-speed film vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (1/3 dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, conventional F-speed bitewing radiography had a summary ROC AUC of 0.98 for proximal caries detection at the 1/3 dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding

Visual examination (VE) vs Histological validation of dental tissues (gold standard)

Diagnostic accuracy for proximal caries detection (1/3 dentin detection level) · Extracted permanent posterior teeth with proximal surfaces (in vitro)

Under in vitro conditions, visual examination had a pooled specificity of 0.84 (95% CI 0.76, 0.89) for proximal caries detection at the 1/3 dentin detection level; ungraded.

4 findings from 1 review
Show 1 more finding