Non-Cochrane · DTA · PM34480645

Systematic review and meta-analysis of diagnostic studies of proximal surface caries.

Janjic Rankovic Mila, Kapor Svetlana, Khazaei Yegane, et al. · Clinical oral investigations · 2021 · PubMed · doi:10.1007/s00784-021-04113-1

Oral HealthDental cariesAMSTAR 2: not applicable
Full textEvery question was extracted from the full text, figures and supplements by two independent LLM extractors, disagreements adjudicated. Rows cite where in the review the value comes from.
Questions
93
current version
GRADE rated
0
No studies
1
Included studies
36
linked below
No studies 1 · 1.1%Ungraded 92 · 98.9%Very low 0 · 0%Low 0 · 0%Moderate 0 · 0%High 0 · 0%

Ungraded means an effect was reported without a GRADE rating; it is not a gap. No studies means the review found nothing to answer the question. The six categories are a display order, not a score.

Review question

ElementPlanned (methods)Found (results)
PopulationIn vivo and in vitro diagnostic studies of primary caries on proximal surfaces of permanent posterior teeth; primary teeth, restorations, secondary caries, and artificially induced lesions excluded36 studies pooled (n=31 in vitro, n=5 in vivo); 129 met initial inclusion, 43 low/moderate RoB, 36 with acceptable data reporting
InterventionCaries detection methods: visual examination (with/without tactile); conventional bitewing radiography (D/E/F-speed film); digital bitewing radiography (sensor, phosphor plate); laser fluorescence (DIAGNOdent 2095 and 2190/pen); fibre-optic transillumination (FOTI)Poolable data for VE, conventional BWR (D/E/F-speed), digital BWR (sensor, phosphor plate), and laser fluorescence (DIAGNOdent 2190); FOTI yielded no poolable studies; DIAGNOdent 2095 not pooled
ComparisonReference/gold standard: histological validation (in vitro); direct visual examination after tooth separation or bioptical cavity preparation (in vivo)Histology (in vitro); tooth separation / bioptical cavity preparation or direct VE (in vivo)
OutcomesDiagnostic test accuracy: sensitivity, specificity, Az/AUC from ROC curves, and/or reliability/reproducibility (Kappa)Pooled sensitivity, specificity, diagnostic odds ratio (DOR), and AUC by method, threshold, and setting (Table 3); clinical/in vivo data limited
Study designsIn vitro (laboratory) and in vivo (clinical) settingsIn vitro (n=31) predominated; in vivo (n=5) limited
TimingNot specifiedNot specified

Search date: Up to 31 December 2018 (English-language publications). Databases: MEDLINE (PubMed); EMBASE

Questions and results

Visual examination (VE) vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
12 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, visual examination had a pooled sensitivity of 0.64 (95% CI 0.42, 0.81) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.64 [0.42, 0.81]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, visual examination had a summary ROC AUC of 0.84 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.849 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, visual examination had a summary ROC AUC of 0.85 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.851 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, visual examination had a pooled diagnostic odds ratio of 11.49 (95% CI 5.19, 25.46) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 11.49 [5.19, 25.46]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, visual examination had a pooled specificity of 0.85 (95% CI 0.74, 0.92) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.85 [0.74, 0.92]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, visual examination had a pooled specificity of 0.84 (95% CI 0.76, 0.89) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.84 [0.76, 0.89]1 studyUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, visual examination had a summary ROC AUC of 0.95 for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.951 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, visual examination had a pooled diagnostic odds ratio of 13.26 (95% CI 1.44, 122.51) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 13.26 [1.44, 122.51]1 studyUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, visual examination had a pooled sensitivity of 0.93 (95% CI 0.77, 0.98) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.93 [0.77, 0.98]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, visual examination had a pooled specificity of 0.99 (95% CI 0.94, 0.999) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.99 [0.94, 0.999]1 studyUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, visual examination had a pooled diagnostic odds ratio of 67.0 (95% CI 14.7, 304.9) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 67.0 [14.7, 304.9]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, visual examination had a pooled sensitivity of 0.09 (95% CI 0.04, 0.24) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.09 [0.04, 0.24]1 studyUngraded

Visual examination (VE) vs Direct visual examination after tooth separation or bioptical cavity preparation

Population: Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)
4 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, visual examination had a pooled specificity of 0.76 (95% CI 0.11, 0.99) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.76 [0.11, 0.99]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, visual examination had a pooled diagnostic odds ratio of 1.84 (95% CI 0.03, 103.47) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 1.84 [0.03, 103.47]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, visual examination had a summary ROC AUC of 0.53 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.532 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, visual examination had a pooled sensitivity of 0.32 (95% CI 0.07, 0.74) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.32 [0.07, 0.74]2 studiesUngraded

Conventional bitewing radiography, D-speed film vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled diagnostic odds ratio of 31.0 (95% CI 3.3, 291.3) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 31.0 [3.3, 291.3]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled diagnostic odds ratio of 5.2 (95% CI 1.6, 16.8) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 5.2 [1.6, 16.8]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled specificity of 0.99 (95% CI 0.96, 0.999) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.99 [0.96, 0.999]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled sensitivity of 0.17 (95% CI 0.07, 0.40) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.17 [0.07, 0.40]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled sensitivity of 0.19 (95% CI 0.13, 0.28) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.19 [0.13, 0.28]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a summary ROC AUC of 0.55 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.552 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a pooled specificity of 0.96 (95% CI 0.91, 0.99) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.96 [0.91, 0.99]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional D-speed bitewing radiography had a summary ROC AUC of 0.91 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.911 studyUngraded

Conventional bitewing radiography, E-speed film vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled sensitivity of 0.42 (95% CI 0.28, 0.57) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.42 [0.28, 0.57]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a summary ROC AUC of 0.82 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.829 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled specificity of 0.90 (95% CI 0.84, 0.93) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.90 [0.84, 0.93]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled specificity of 0.94 (95% CI 0.71, 0.99) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.94 [0.71, 0.99]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled diagnostic odds ratio of 6.80 (95% CI 4.58, 10.10) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 6.80 [4.58, 10.10]9 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled diagnostic odds ratio of 31.40 (95% CI 14.31, 68.93) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 31.40 [14.31, 68.93]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a summary ROC AUC of 0.92 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.925 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional E-speed bitewing radiography had a pooled sensitivity of 0.67 (95% CI 0.19, 0.95) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.67 [0.19, 0.95]5 studiesUngraded

Conventional bitewing radiography, E-speed film vs Direct visual examination after tooth separation or bioptical cavity preparation

Population: Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)
4 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, conventional E-speed bitewing radiography had a pooled specificity of 0.91 (95% CI 0.80, 0.96) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.91 [0.80, 0.96]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, conventional E-speed bitewing radiography had a pooled sensitivity of 0.35 (95% CI 0.19, 0.51) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.35 [0.19, 0.51]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, conventional E-speed bitewing radiography had a summary ROC AUC of 0.74 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.741 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, conventional E-speed bitewing radiography had a pooled diagnostic odds ratio of 4.8 (95% CI 1.48, 15.48) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 4.8 [1.48, 15.48]1 studyUngraded

Conventional bitewing radiography, F-speed film vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
12 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled specificity of 0.88 (95% CI 0.74, 0.95) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.88 [0.74, 0.95]8 studiesUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled sensitivity of 0.54 (95% CI 0.36, 0.70) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.54 [0.36, 0.70]1 studyUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled specificity of 0.996 (95% CI 0.96, 1) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.996 [0.96, 1]1 studyUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled diagnostic odds ratio of 283.6 (95% CI 16.1, 5010.0) for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 283.6 [16.1, 5010.0]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled diagnostic odds ratio of 8.7 (95% CI 4.2, 18.1) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 8.7 [4.2, 18.1]3 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled sensitivity of 0.42 (95% CI 0.29, 0.56) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.42 [0.29, 0.56]3 studiesUngraded
Diagnostic accuracy for proximal caries detection (1/3 dentin detection level)
1/3 dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a summary ROC AUC of 0.98 for proximal caries detection at the 1/3 dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.981 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a summary ROC AUC of 0.72 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.728 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a summary ROC AUC of 0.81 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.813 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled diagnostic odds ratio of 5.34 (95% CI 2.55, 11.17) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 5.34 [2.55, 11.17]8 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled specificity of 0.92 (95% CI 0.86, 0.96) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.92 [0.86, 0.96]3 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, conventional F-speed bitewing radiography had a pooled sensitivity of 0.43 (95% CI 0.31, 0.57) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.43 [0.31, 0.57]8 studiesUngraded

Digital bitewing radiography, sensor-based (CCD/CMOS) vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled diagnostic odds ratio of 12.84 (95% CI 6.56, 25.13) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 12.84 [6.56, 25.13]4 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled sensitivity of 0.35 (95% CI 0.26, 0.45) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.35 [0.26, 0.45]12 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled specificity of 0.90 (95% CI 0.85, 0.93) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.90 [0.85, 0.93]12 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled diagnostic odds ratio of 5.01 (95% CI 2.83, 8.88) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 5.01 [2.83, 8.88]12 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled specificity of 0.95 (95% CI 0.93, 0.97) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.95 [0.93, 0.97]4 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a summary ROC AUC of 0.90 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.904 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a pooled sensitivity of 0.36 (95% CI 0.30, 0.42) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.36 [0.30, 0.42]4 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, sensor-based digital bitewing radiography had a summary ROC AUC of 0.74 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.7412 studiesUngraded

Digital bitewing radiography, sensor-based (CCD/CMOS) vs Direct visual examination after tooth separation or bioptical cavity preparation

Population: Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled specificity of 0.50 (95% CI 0.02, 0.98) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.50 [0.02, 0.98]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled specificity of 0.93 (95% CI 0.77, 0.98) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.93 [0.77, 0.98]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled sensitivity of 0.96 (95% CI 0.91, 0.98) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.96 [0.91, 0.98]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled sensitivity of 0.55 (95% CI 0.42, 0.67) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.55 [0.42, 0.67]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled diagnostic odds ratio of 17.11 (95% CI 3.73, 78.39) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 17.11 [3.73, 78.39]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a summary ROC AUC of 0.89 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.891 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a pooled diagnostic odds ratio of 24.4 (95% CI 0.44, 1359.98) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 24.4 [0.44, 1359.98]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, sensor-based digital bitewing radiography had a summary ROC AUC of 0.82 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.821 studyUngraded

Digital bitewing radiography, phosphor plate (PSP) vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a summary ROC AUC of 0.82 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.8211 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a summary ROC AUC of 0.92 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.922 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled sensitivity of 0.41 (95% CI 0.25, 0.61) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.41 [0.25, 0.61]11 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled specificity of 0.89 (95% CI 0.83, 0.93) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.89 [0.83, 0.93]11 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled sensitivity of 0.86 (95% CI 0.03, 0.99) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.86 [0.03, 0.99]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled diagnostic odds ratio of 26.06 (95% CI 7.29, 93.21) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 26.06 [7.29, 93.21]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled specificity of 0.86 (95% CI 0.11, 0.99) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.86 [0.11, 0.99]2 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, phosphor-plate digital bitewing radiography had a pooled diagnostic odds ratio of 5.56 (95% CI 3.04, 10.16) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 5.56 [3.04, 10.16]11 studiesUngraded

Digital bitewing radiography, phosphor plate (PSP) vs Direct visual examination after tooth separation or bioptical cavity preparation

Population: Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)
4 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, phosphor-plate digital bitewing radiography had a pooled specificity of 0.60 (95% CI 0.20, 0.90) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.60 [0.20, 0.90]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, phosphor-plate digital bitewing radiography had a pooled sensitivity of 0.83 (95% CI 0.77, 0.88) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.83 [0.77, 0.88]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, phosphor-plate digital bitewing radiography had a pooled diagnostic odds ratio of 7.5 (95% CI 1.19, 47.13) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 7.5 [1.19, 47.13]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, phosphor-plate digital bitewing radiography had a summary ROC AUC of 0.79 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.791 studyUngraded

Laser fluorescence (DIAGNOdent 2190/pen) vs Histological validation of dental tissues (gold standard)

Population: Extracted permanent posterior teeth with proximal surfaces (in vitro)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a summary ROC AUC of 0.83 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.835 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled specificity of 0.89 (95% CI 0.76, 0.95) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.89 [0.76, 0.95]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled diagnostic odds ratio of 30.79 (95% CI 8.74, 108.51) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 30.79 [8.74, 108.51]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled sensitivity of 0.79 (95% CI 0.62, 0.90) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.79 [0.62, 0.90]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled specificity of 0.81 (95% CI 0.78, 0.85) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.81 [0.78, 0.85]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a summary ROC AUC of 0.91 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.915 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled sensitivity of 0.82 (95% CI 0.58, 0.94) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.82 [0.58, 0.94]5 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vitro conditions, laser fluorescence (DIAGNOdent 2190) had a pooled diagnostic odds ratio of 23.09 (95% CI 7.01, 76.04) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 23.09 [7.01, 76.04]5 studiesUngraded

Laser fluorescence (DIAGNOdent 2190/pen) vs Direct visual examination after tooth separation or bioptical cavity preparation

Population: Patients with permanent posterior teeth, proximal surfaces assessed clinically (in vivo)
8 questions
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled specificity of 0.60 (95% CI 0.15, 0.93) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.60 [0.15, 0.93]3 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled sensitivity of 0.92 (95% CI 0.82, 0.96) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.92 [0.82, 0.96]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a summary ROC AUC of 0.96 for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.961 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a summary ROC AUC of 0.63 for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
AUC 0.633 studiesUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled diagnostic odds ratio of 99 (95% CI 22.01, 445.30) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 99 [22.01, 445.30]1 studyUngraded
Diagnostic accuracy for proximal caries detection (caries detection level)
Caries detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled specificity of 0.90 (95% CI 0.73, 0.97) for proximal caries detection at the caries detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Specificity 0.90 [0.73, 0.97]1 studyUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled diagnostic odds ratio of 2.18 (95% CI 0.24, 19.95) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
DOR 2.18 [0.24, 19.95]3 studiesUngraded
Diagnostic accuracy for proximal caries detection (dentin detection level)
Dentin detection level
Under in vivo conditions, laser fluorescence (DIAGNOdent 2190) had a pooled sensitivity of 0.63 (95% CI 0.58, 0.68) for proximal caries detection at the dentin detection level; ungraded.
Source: Table 3 (bivariate diagnostic random-effects meta-analysis); Results · same question in other reviews
Sensitivity 0.63 [0.58, 0.68]3 studiesUngraded

Fibre-optic transillumination (FOTI) vs Histological validation (in vitro) or direct visual examination after tooth separation / bioptical cavity preparation (in vivo)

Population: Permanent posterior teeth, proximal surfaces (in vitro extracted teeth and in vivo patients)
1 question
OutcomeEffectEvidenceStatus / GRADE
Diagnostic accuracy for proximal caries detection
For fibre-optic transillumination (FOTI), studies were identified (in vitro N=4, in vivo N=3) but none had sufficiently low risk of bias and adequate reporting to be pooled; no meta-analytic estimate could be derived.
Source: Table 2 (FOTI: in vitro N=4, in vivo N=3, none passed quality assessment); Table 3 (FOTI row, no estimates); Results · same question in other reviews
no estimate extracted0 studiesNo studies

Included primary studies 36

Open as a trial list
StudyAlso included byOutcomes it informs hereTrial PICOResults
Abesi 2012
Diagnostic accuracy of digital and conventional radiography in the detection of non-cavitated approximal dental caries.
PMID:23329955
this review onlynot statednot extracted yetnot extracted yet
Abogazalah 2019
In vitro visual and visible light transillumination methods for detection of natural non-cavitated approximal caries
DOI:10.1007/s00784-018-2546-3
this review onlynot statednot extracted yetnot extracted yet
Ariji 1998
matched by author and year (probable)
this review onlynot statednot extracted yetnot extracted yet
Astvaldsdottir 2012
Approximal Caries Detection by DIFOTI: In Vitro Comparison of Diagnostic Accuracy/Efficacy with Film and Digital Radiography.
PMID:23213335
this review onlynot statednot extracted yetnot extracted yet
Behere 2011
Reliability of Logicon caries detector in the detection and depth assessment of dental caries: An in-vitro study
DOI:10.4103/0970-9290.84277
this review onlynot statednot extracted yetnot extracted yet
Bozdemir 2016
Comparison of different caries detectors for approximal caries detection.
PMID:30894987
this review onlynot statednot extracted yetnot extracted yet
da Silva Neto 2008
Radiographic diagnosis of incipient proximal caries: an ex-vivo study
DOI:10.1590/s0103-64402008000200002
this review onlynot statednot extracted yetnot extracted yet
de Souza 2014
In vitro performance of a pen-type laser fluorescence device and bitewing radiographs for approximal caries detection in permanent and…
DOI:10.4103/0970-9290.152165
this review onlynot statednot extracted yetnot extracted yet
Ekstrand 2011
The Reliability and Accuracy of Two Methods for Proximal Caries Detection and Depth on Directly Visible Proximal Surfaces: An in vitro Study
DOI:10.1159/000324439
this review onlynot statednot extracted yetnot extracted yet
Forner-Navarro 2008
matched by author and year (probable)
this review onlynot statednot extracted yetnot extracted yet
Haak 2003
Influence of displayed image size on radiographic detection of approximal caries
DOI:10.1259/dmfr/17654484
this review onlynot statednot extracted yetnot extracted yet
Haiter-Neto 2009
Task-specific enhancement filters in storage phosphor images from the Vistascan system for detection of proximal caries lesions of known…
DOI:10.1016/j.tripleo.2008.09.031
this review onlynot statednot extracted yetnot extracted yet
Hintze 2003
Diagnostic Outcome of Methods Frequently Used for Caries Validation
DOI:10.1159/000069016
this review onlynot statednot extracted yetnot extracted yet
Huth 2010
In vivo performance of a laser fluorescence device for the approximal detection of caries in permanent molars
DOI:10.1016/j.jdent.2010.09.001
this review onlynot statednot extracted yetnot extracted yet
Isidor 2009
Effect of monitor display on detection of approximal caries lesions in digital radiographs
DOI:10.1259/dmfr/21071028
this review onlynot statednot extracted yetnot extracted yet
Jan 2016
Proximal caries lesion detection using the <scp>C</scp>anary <scp>C</scp>aries <scp>D</scp>etection <scp>S</scp>ystem: an <i>in vitro</i>…
DOI:10.1111/jicd.12163
this review onlynot statednot extracted yetnot extracted yet
Kuhnisch 2016
In vivo validation of near-infrared light transillumination for interproximal dentin caries detection
DOI:10.1007/s00784-015-1559-4
this review onlynot statednot extracted yetnot extracted yet
Lussi 2006
Detection of Approximal Caries with a New Laser Fluorescence Device
DOI:10.1159/000091054
this review onlynot statednot extracted yetnot extracted yet
Matalon 2003
Diagnosis of approximal caries: Bite-wing radiology versus the Ultrasound Caries Detector. An in vitro study
DOI:10.1067/moe.2003.164
this review onlynot statednot extracted yetnot extracted yet
Menem 2017
The diagnostic accuracy of a laser fluorescence device and digital radiography in detecting approximal caries lesions in posterior…
PMID:28194533
this review onlynot statednot extracted yetnot extracted yet
Mitropoulos 2010
Diagnostic performance of the visual caries classification system ICDAS II versus radiography and micro-computed tomography for proximal…
DOI:10.1016/j.jdent.2010.07.005
this review onlynot statednot extracted yetnot extracted yet
Neuhaus 2015
Diagnostic performance of a new red light LED device for approximal caries detection
DOI:10.1007/s10103-014-1607-3
this review onlynot statednot extracted yetnot extracted yet
Ozkan 2017
Clinical evaluation of near-infrared light transillumination in approximal dentin caries detection
DOI:10.1007/s10103-017-2265-z
this review onlynot statednot extracted yetnot extracted yet
Pontual 2010
Comparison of digital systems and conventional dental film for the detection of approximal enamel caries.
PMID:20841461
this review onlynot statednot extracted yetnot extracted yet
Ricketts 1997
An evaluation of the diagnostic yield from bitewing radiographs of small approximal and occlusal carious lesions in a low prevalence…
DOI:10.1038/sj.bdj.4809298
this review onlynot statednot extracted yetnot extracted yet
Rockenbach 2008
matched by author and year (probable)
this review onlynot statednot extracted yetnot extracted yet
Safi 2015
matched by author and year (probable)
this review onlynot statednot extracted yetnot extracted yet
Senel 2010
Diagnostic accuracy of different imaging modalities in detection of proximal caries.
PMID:21062944
this review onlynot statednot extracted yetnot extracted yet
Shimada 2014
Noninvasive cross‐sectional imaging of proximal caries using swept‐source optical coherence tomography (SS‐OCT) <i>in vivo</i>
DOI:10.1002/jbio.201200210
this review onlynot statednot extracted yetnot extracted yet
Tonkaboni 2019
Comparison of diagnostic effects of infrared imaging and bitewing radiography in proximal caries of permanent teeth
DOI:10.1007/s10103-018-2663-x
this review onlynot statednot extracted yetnot extracted yet
Vivek 2015
Comparison of Diagnostic Ability of Storage Phosphor Plate in Detecting Proximal Caries with Direct Measurement by Stereomicroscope: A…
PMID:26664712
this review onlynot statednot extracted yetnot extracted yet
Wenzel 2002
Accuracy of computer‐automated caries detection in digital radiographs compared with human observers
DOI:10.1034/j.1600-0447.2002.21245.x
this review onlynot statednot extracted yetnot extracted yet
Wenzel 2007a
Risk Factors for a False Positive Test Outcome in Diagnosis of Caries in Approximal Surfaces: Impact of Radiographic Modality and Observer…
DOI:10.1159/000099314
this review onlynot statednot extracted yetnot extracted yet
Wenzel 2007b
Influence of spatial resolution and bit depth on detection of small caries lesions with digital receptors
DOI:10.1016/j.tripleo.2006.05.016
this review onlynot statednot extracted yetnot extracted yet
ZangooeiBooshehry 2010
Efficacy of application of pseudocolor filters in the detection of interproximal caries.
PMID:22991603
this review onlynot statednot extracted yetnot extracted yet
Zayet 2014
Effect of changing the kilovoltage peak on radiographic caries assessment in digital and conventional radiography.
PMID:25279340
this review onlynot statednot extracted yetnot extracted yet