Non-Cochrane · intervention · PM42676689

Comparison of the Effectiveness of the C-MAC Versus the McGrath Videolaryngoscope for Adult Endotracheal Intubation in Clinical Patients, Simulated Patients, and Manikin Settings: A Systematic Review and Meta-analysis.

Dimou Evgenia, Aretha Diamanto, Beskou-Kontou Theano, et al. · Anesthesiology research and practice · 2026 · PubMed · doi:10.1155/anrp/1575867

Findings

Population Adult clinical patients undergoing endotracheal intubation

8 findings

McGrath videolaryngoscopeC-MAC videolaryngoscopeFirst-pass intubation success
Adult clinical patient studies · Random-effects model; Excluding studies with intubation time as primary outcome
OR 0.89 [0.45, 1.78]
GRADE not reported
3 studies · 530 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeTime to intubation
Adult clinical patient studies · Random-effects model
MD 9.59 [-1.35, 20.52]
Very low
Downgraded: imprecision, inconsistency, publication bias
6 studies · 780 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeFirst-pass intubation success
Adult clinical patient studies · Fixed-effect model
OR 0.54 [0.36, 0.82]
GRADE not reported
6 studies · 780 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeTime to intubation
Adult clinical patient studies · Fixed-effect model; Excluding studies with first-pass success as primary outcome
MD 14.12 [10.10, 18.14]
GRADE not reported
4 studies · 330 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeTime to intubation
Adult clinical patient studies · Random-effects model; Excluding studies with first-pass success as primary outcome
MD 14.29 [3.81, 24.76]
GRADE not reported
4 studies · 330 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeFirst-pass intubation success
Adult clinical patient studies · Random-effects model
OR 0.54 [0.29, 1.01]
Low
Downgraded: imprecision, inconsistency
6 studies · 780 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeTime to intubation
Adult clinical patient studies · Fixed-effect model
MD 7.40 [4.54, 10.26]
GRADE not reported
6 studies · 780 participants
McGrath videolaryngoscopeC-MAC videolaryngoscopeFirst-pass intubation success
Adult clinical patient studies · Fixed-effect model; Excluding studies with intubation time as primary outcome
OR 0.83 [0.49, 1.42]
GRADE not reported
3 studies · 530 participants

Included primary studies 9

Open in Studies
StudyYearAlso included byOutcomes it informs hereTrial data
Akbas 2019
A Comparison of McGrath MAC Versus C-MAC Videolaryngoscopes in Morbidly Obese Patients Undergoing Bariatric Surgery: A Randomized,…
DOI:10.1089/bari.2018.0052
2019this review onlyFirst-pass intubation success; Time to intubationnot extracted yet
Gupta 2022
Comparison of C-MAC and McGrathMAC Videolaryngoscopes for Intubation in Patients with Normal Airway by Donned Anaesthesiologists Using an…
PMID:35979971
2022this review onlyFirst-pass intubation success; Time to intubationnot extracted yet
Kleine-Brueggeney 2016
Evaluation of six videolaryngoscopes in 720 patients with a simulated difficult airway: a multicentre randomized controlled trial
DOI:10.1093/bja/aew058
2016this review onlyFirst-pass intubation success; Time to intubationnot extracted yet
Ng 2012
Randomized controlled trial comparing the McGrath videolaryngoscope with the C-MAC videolaryngoscope in intubating adult patients with…
DOI:10.1093/bja/aes145
2012this review onlyFirst-pass intubation success; Time to intubationnot extracted yet
Pieters 2016
Comparison of seven videolaryngoscopes with the Macintosh laryngoscope in manikins by experienced and novice personnel
DOI:10.1111/anae.13413
2016this review onlynot statednot extracted yet
Shin 2016
Comparing McGRATH® MAC, C-MAC®, and Macintosh Laryngoscopes Operated by Medical Students: A Randomized, Crossover, Manikin Study
DOI:10.1155/2016/8943931
2016this review onlynot statednot extracted yet
Vig 2021
Comparison of Mc Grath-MAC and C-MAC video laryngoscopes for intubation in a COVID simulated mannequin by novice users wearing face…
PMID:34188630
2021this review onlynot statednot extracted yet
Yumul 2016
Comparison of three video laryngoscopy devices to direct laryngoscopy for intubating obese patients: a randomized controlled trial
DOI:10.1016/j.jclinane.2015.12.042
2016this review onlyFirst-pass intubation success; Time to intubationnot extracted yet
Zhang 2021
Comparing the first-attempt tracheal intubation success of the hyperangulated McGrath® X-blade vs the Macintosh-type CMAC…
DOI:10.1007/s10877-021-00746-5
2021this review onlyFirst-pass intubation success; Time to intubationnot extracted yet

Review question

ElementPlanned (methods)Found (results)
PopulationAdult clinical patients undergoing endotracheal intubation, simulated-patient clinical scenarios, and manikin-based adult airway modelsAdult clinical patients (n=780; 6 studies) pooled; manikin studies (3; 322 operators or datasets) summarized qualitatively; obesity (2), COVID-19 intubation box (1), normal or simulated difficult airway (3)
InterventionIntubation with the C-MAC or McGrath videolaryngoscopeC-MAC and McGrath videolaryngoscopes (some studies also had other videolaryngoscopes or Macintosh arms)
ComparisonThe alternative device (McGrath or C-MAC)Direct comparison of C-MAC versus McGrath
OutcomesPrimary: first-pass intubation success, time to intubation; secondary: total intubation success, number of attempts, glottic view, optimization maneuvers, operator difficulty or preference, airway complicationsFirst-pass intubation success (6 studies; n=780); time to intubation (6 studies; n=780); complications summarized narratively
Study designsRCTs, prospective randomized studies, randomized crossover designs9 randomized studies (6 clinical RCTs, 3 manikin studies)
TimingNot specifiedNot specified

AMSTAR 2 items

Two independent extractors rated each item. Their ratings are shown side by side because they have not been reconciled yet. The overall rating follows Shea et al. 2017: one critical flaw gives Low, more than one gives Critically Low.

ItemExtractor AExtractor BAgree
1Research question includes PICO componentsYesYesyes
2Protocol established before the reviewcriticalPartial YesPartial Yesyes
3Choice of study designs explainedNoNoyes
4Comprehensive literature searchcriticalNoPartial Yesno
5Study selection in duplicateYesYesyes
6Data extraction in duplicateYesYesyes
7List of excluded studies with reasonscriticalNoNoyes
8Included studies described in detailPartial YesYesno
9Risk of bias assessed satisfactorilycriticalRCT: YesNRSI: N/A · RCT: Yesno
10Funding of included studies reportedYesYesyes
11Appropriate meta-analysis methodscriticalRCT: YesNRSI: N/A · RCT: Yesno
12Impact of risk of bias on the pooled resultsYesNono
13Risk of bias accounted for in interpretationcriticalYesYesyes
14Heterogeneity explainedYesYesyes
15Publication bias investigatedcriticalYesYesyes
16Conflicts of interest reportedYesYesyes
OverallCritically LowLowno