Cochrane · systematic review · CD003531

Pressure modification or humidification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoea

Kennedy, B, Lasserson, TJ, Wozniak, DR, et al. · Cochrane Database of Systematic Reviews · 2019 · Cochrane Library

AirwaysAMSTAR 2: not assessed
SoF tableOnly the outcomes this review put in its Summary of Findings table are listed (the Cochrane Handbook caps them at about seven per comparison). Its other analyses are not shown here.
Questions
32
current version
GRADE rated
25
No studies
7
Included studies
NR
No studies 7 · 21.9%Ungraded 0 · 0%Very low 4 · 12.5%Low 14 · 43.8%Moderate 5 · 15.6%High 2 · 6.3%

Ungraded means an effect was reported without a GRADE rating; it is not a gap. No studies means the review found nothing to answer the question. The six categories are a display order, not a score.

Questions and results

Auto‐CPAP compared to fixed CPAP for sleep apnoea in adults

Population: adults with a diagnosis of OSA
8 questions
OutcomeEffectEvidenceStatus / GRADE
Average machine usage (hours per night) Median
Auto‐CPAP vs fixed CPAP
no estimate extracted31 studies · 1452 participantsModerate
Downgraded: risk of bias
Number of participants using machine for 4 or more hours per night
Auto‐CPAP vs fixed CPAP
OR 1.16 [0.75, 1.81]2 studies · 346 participantsLow
Downgraded: risk of bias, imprecision
Symptoms
Auto‐CPAP vs fixed CPAP
no estimate extracted25 studies · 1285 participantsModerate
Downgraded: risk of bias
Withdrawals (parallel group trials/first arm cross‐over trials) Median
Auto‐CPAP vs fixed CPAP
OR 0.9 [0.64, 1.27]13 studies · 1275 participantsModerate
Downgraded: imprecision
Quality of life
Auto‐CPAP vs fixed CPAP
no estimate extracted3 studies · 352 participantsLow
Downgraded: risk of bias, imprecision
AHI measured by number of events/hr Median
Auto‐CPAP vs fixed CPAP
no estimate extracted26 studies · 1256 participantsHigh
Blood pressure (mmHg)
Auto‐CPAP vs fixed CPAP
no estimate extracted2 studies · 353 participantsModerate
Downgraded: imprecision
Adverse events (machine tolerability outcomes)
Auto‐CPAP vs fixed CPAP
no estimate extracted9 studies · 574 participantsVery low
Downgraded: risk of bias, imprecision, inconsistency

Bi‐PAP compared to fixed CPAP for sleep apnoea in adults

Population: adults with a diagnosis of sleep apnoea
8 questions
OutcomeEffectEvidenceStatus / GRADE
Machine usage (hours/night)
Bi‐PAP vs fixed CPAP
no estimate extracted4 studies · 268 participantsLow
Downgraded: risk of bias, imprecision
Machine usage assessed by number of participants using machine for 4 or more hours per night ‐ not measured
Bi‐PAP vs fixed CPAP
no estimate extractedNRNo studies
Symptoms
Bi‐PAP vs fixed CPAP
no estimate extracted4 studies · 226 participantsLow
Downgraded: risk of bias, imprecision
Withdrawals (parallel group trials/first arm cross‐over trials)
Bi‐PAP vs fixed CPAP
OR 0.55 [0.26, 1.17]3 studies · 261 participantsLow
Downgraded: risk of bias, imprecision, inconsistency
Quality of life
Bi‐PAP vs fixed CPAP
no estimate extracted1 study · 151 participantsLow
Downgraded: risk of bias, imprecision, inconsistency
AHI
Bi‐PAP vs fixed CPAP
no estimate extracted2 studies · 179 participantsLow
Downgraded: risk of bias, imprecision, inconsistency
Blood pressure ‐ not measured
Bi‐PAP vs fixed CPAP
no estimate extractedNRNo studies
Adverse events (machine tolerability outcomes)
Bi‐PAP vs fixed CPAP
no estimate extracted5 studies · 239 participantsVery low
Downgraded: risk of bias, imprecision

CPAP with expiratory pressure relief compared to fixed CPAP for adults with obstructive sleep apnoea

Population: adults with a diagnosis of sleep apnoea
8 questions
OutcomeEffectEvidenceStatus / GRADE
Machine usage assessed by hours/night
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extracted9 studies · 609 participantsLow
Downgraded: risk of bias
Machine usage assessed by number of participants using machine for 4 or more hours per night ‐ not measured
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extractedNRNo studies
Symptoms
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extracted6 studies · 515 participantsModerate
Downgraded: risk of bias
Withdrawals (parallel group trials/first arm cross‐over trials)
CPAP with expiratory pressure relief vs fixed CPAP
OR 0.86 [0.48, 1.55]2 studies · 298 participantsLow
Downgraded: risk of bias, imprecision, inconsistency
Quality of life
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extracted1 study · 74 participantsLow
Downgraded: risk of bias, imprecision, inconsistency
AHI measured by number of events/hr
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extracted5 studies · 342 participantsHigh
Blood pressure ‐ not measured
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extractedNRNo studies
Adverse events (machine tolerability outcomes)
CPAP with expiratory pressure relief vs fixed CPAP
no estimate extracted6 studies · 577 participantsVery low
Downgraded: risk of bias, imprecision, inconsistency

Heated humidification + fixed CPAP compared to fixed CPAP alone in adults with obstructive sleep apnoea

Population: adults with a diagnosis of sleep apnoea
8 questions
OutcomeEffectEvidenceStatus / GRADE
Machine usage assessed by hours per night
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extracted6 studies · 277 participantsLow
Downgraded: risk of bias
Machine usage assessed by number of participants using machine for 4 or more hours per night ‐ not measured
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extractedNRNo studies
Symptoms
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extracted4 studies · 184 participantsLow
Downgraded: risk of bias, imprecision
Withdrawals (parallel group trials/first arm cross‐over trials)
Heated humidification + fixed CPAP vs fixed CPAP alone
OR 1 [0.45, 2.24]3 studies · 316 participantsVery low
Downgraded: risk of bias, imprecision
Quality of life
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extractedNRNo studies
AHI measured by number of events/hr
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extracted1 study · 44 participantsLow
Downgraded: risk of bias, imprecision
Blood pressure ‐ not measured
Heated humidification + fixed CPAP vs fixed CPAP alone
no estimate extractedNRNo studies
Adverse events (machine tolerability outcomes) assessed as number of participants experiencing a blocked nose
Heated humidification + fixed CPAP vs fixed CPAP alone
OR 0.32 [0.16, 0.63]2 studies · 147 participantsLow
Downgraded: risk of bias, imprecision